APPLIED OPERATIONS
/Image Acquisition
2-48
Page
2.
While observing the image in the [Live] panel, locate the upper edge of the range to
be observed by moving down the stage using the <Z stage coarse adjustment> and
<Z stage fine adjustment> buttons in the [Z Stage] sub-panel.
When using the FLUOVIEW system with an inverted microscope, locate the bottom
edge of the range to be observed by moving down the revolving nosepiece using
the <Z stage coarse adjustment> and <Z stage fine adjustment> buttons.
3. When the upper edge position is located, click the <Set> button. The [Start Z] text
box will show the scan start position of the range of the multiple sections to be
observed (Z-direction scanning range).
4. While observing the image in the [Live] panel, locate the bottom edge of the range
to be observed by moving up the stage using the <Z stage coarse adjustment> and
<Z stage fine adjustment> buttons in the [Z Stage] sub-panel.
When using the FLUOVIEW system with an inverted microscope, locate the upper
edge of the range to be observed by moving up the revolving nosepiece using the
<Z stage coarse adjustment> and <Z stage fine adjustment> buttons.
5.
When the position is located, click the <Set> button. The [Stop Z] text box will show
the scanning stop position of the range of the multiple sections to be observed (Z-
direction scanning range).
Summary of Contents for FLUOVIEW FV500
Page 2: ......
Page 8: ......
Page 10: ......
Page 24: ......
Page 110: ......
Page 327: ...APPLIED OPERATIONS Viewing 3D Image 2 217 Page Fig 2 106 Panel Showing Stereo 3D Images ...
Page 396: ......
Page 402: ......
Page 414: ......
Page 420: ......
Page 426: ......
Page 428: ......
Page 430: ......
Page 458: ......
Page 460: ......
Page 462: ......
Page 468: ......
Page 469: ......