APPLIED OPERATIONS
/Image Acquisition
2-68
Page
z
With types of dyeing of the specimen to observe, the Barrier Filters will be set
automatically to the optical path.
To change the Barrier Filter types, see section 1-3-2-4,”Configuring the Filters “ and
change by [Optical system Configuration] window.
z
Turn off the transmitted light bulb of the microscope.
1. From the page tabs on the bottom right of the [Acquire] panel, select the
[Optics] sub-panel.
Fig. 2-23 [Optics] Sub-panel
2. Click the <Trans. Lamp> button to turn the lamp off (set the button to the
non-pressed status).
z
In the [Acquire] panel, make sure that the [Transmitted] check box in the [Ch5]
group box is check-marked to indicate that Ch3 is ready for image acquisition.
<Trans. Lamp> button
Turns off the transmitted light bulb of the
microscope.
(The lamp lights when the button is in
the pressed status.)
Ch5 check box