APPLIED OPERATIONS
/Image Acquisition
2-72
Page
2-2-6 Acquiring Finer Image (Sequential Scan)
An image without containing superimposition of fluorescence can be obtained by
acquiring the image slice of each laser excitation or channel sequentially.
With this image capturing method, the image of a multiple-dyed specimen can be
obtained by sequentially acquiring image slice of each type of fluorescence.
1. Select the [Surface XY-Note] option button in the [Scan Mode] group box in the
[Acquire] panel, then select <Sequen> in the list shown below.
Selecting <Sq Clp>, <Sq Rct>, or <Line Sq> enables the image acquisition of the
desired rectangle area or that of the rectangle area at desired angle, or a line
sequential scan..
Line Sequential Scan
Ordinary sequential scan is performed for every one frame for every group,
whereas line sequential scan is performed for every one line for every group.
Therefore, line sequential scan is of advantage in minimizing the time lag
between wavelength and acquiring a fewer cross-talk image.
Image acquisition in the line mode can be performed when
you use the FV300 system with AOTF (FV5-COMBA).
2.
Select the observation mode with the option buttons in the [Acquire] panel.
NOTE
Summary of Contents for FLUOVIEW FV500
Page 2: ......
Page 8: ......
Page 10: ......
Page 24: ......
Page 110: ......
Page 327: ...APPLIED OPERATIONS Viewing 3D Image 2 217 Page Fig 2 106 Panel Showing Stereo 3D Images ...
Page 396: ......
Page 402: ......
Page 414: ......
Page 420: ......
Page 426: ......
Page 428: ......
Page 430: ......
Page 458: ......
Page 460: ......
Page 462: ......
Page 468: ......
Page 469: ......