APPLIED OPERATIONS
/Image Acquisition
2-78
Page
6. The inclination angle of the line can be varied according to the mouse pointer
movement.
Place the moue pointer on an inclining position and click the mouse left button to fix
the line inclination angle.
7.
Acquire the image with the buttons such as the <XYZ>, <XYT>, <XYZT> buttons.
8.
After acquiring the image, select <Normal> under the [Surface XY] option button in
the [Scan Mode] group box in the [Acquire] panel to return from the image
acquisition mode.
2-2-9 Display the change of image intensity (Point Scan)
The change of fluorescence intensity according to time-lapse can be displayed
graphically by irradiating the laser to certain point on the image.
If you have FLUOVIEW TIME COURSE software, image acquisition can be started with
buttons, keys or input of external trigger signals.
1. Acquire an image in the XY observation mode. For the operation procedure, see
section 2-2-1, “Image Acquisition with XY observation”.
2. Select <Point> under the [Point] option button in the [Scan Mode] group box in the
[Acquire] panel.
Shows the X, Y
coordinates on the
image.