APPLIED OPERATIONS
/Image Acquisition
2-98
Page
7 Acquiring Data
The experiment data before and after photobleaching of certain region can be obtained
after performing an ordinary image acquisition in the first scan, an image acquisition in
the Bleach mode in the second scan , and an ordinary image acquisition in the third scan.
When the REX mode is selected in the third scan, the data of the region other than the
specified region can not be obtained since the laser is irradiated only to the specified
region. However, the REX mode is effective to prevent fading of the obtained data.
1. Select the [Disabled] option button in the [AOTF] group box in the [Lasers] sub-
panel.
2. Click the <XYT> button to start image acquisition.
3. Select the [Bleach mode] option button in the [AOTF] group box in the [Lasers]
sub-panel.
Select the [Bleach mode] option button within 10 seconds before the
next scanning is started.
In the TIME COURSE software (optional), when the REX mask file in
the REX or Bleach mode is changed while image scanning is
performed obtaining the real time graph, the laser is irradiated while
the pop-up menu is displayed. however, the real time graph is not
plotted. Please be aware this before using.
<XYT> button
[Disabled] option button
[Bleach mode] option button
NOTE
NOTE
Summary of Contents for FLUOVIEW FV500
Page 2: ......
Page 8: ......
Page 10: ......
Page 24: ......
Page 110: ......
Page 327: ...APPLIED OPERATIONS Viewing 3D Image 2 217 Page Fig 2 106 Panel Showing Stereo 3D Images ...
Page 396: ......
Page 402: ......
Page 414: ......
Page 420: ......
Page 426: ......
Page 428: ......
Page 430: ......
Page 458: ......
Page 460: ......
Page 462: ......
Page 468: ......
Page 469: ......