Getting Started FLUOVIEW/
Basic Operations
1-9
Page
With transmitted light differential interference observation
using an immersion objective, set the microscope’s field
diaphragm so that it inscribes the field of view. Otherwise
the contrast may degrade. (This applies to both visual
observation and laser differential interference observation.)
(5) Condenser, polarizing plate
Condenser for transmitted lighting.
In addition, the rotary turret for the transmitted light DIC slider and the polarizing plate
for differential interference observation (analyzer) are also provided.
•
To perform differential interference observation, engage the transmitted light DIC
slider (optional) matching the objective in use in the light path (For both visual
observation and laser differential interference observation).
•
To perform visual differential interference observation or laser differential
interference observation, engage the polarizing plate in the light path.
(6) Filters
These filters are used to adjust transmitted light.
•
For transmitted observation using laser, disengage the LBD filter from the light
path and engage the FR (frost) filter in the light path by operating the filter levers.
If the FR filter is disengaged from the light path, the image may suffer from stripe
interference.
(7) Intermediate magnification knob
Select the intermediate magnification from 1X or 1.5X.
•
Always set to 1X for using the microscope as a LSM. It cannot be used as a LSM
if this knob is set to 1.5X.
NOTE
Summary of Contents for FLUOVIEW FV500
Page 2: ......
Page 8: ......
Page 10: ......
Page 24: ......
Page 110: ......
Page 327: ...APPLIED OPERATIONS Viewing 3D Image 2 217 Page Fig 2 106 Panel Showing Stereo 3D Images ...
Page 396: ......
Page 402: ......
Page 414: ......
Page 420: ......
Page 426: ......
Page 428: ......
Page 430: ......
Page 458: ......
Page 460: ......
Page 462: ......
Page 468: ......
Page 469: ......