TROUBLESHOOTING GUIDE
1-5
Page
Phenomenon
Cause
Treatment
Manual Ref. Pages
18. The mouse pointer in
the screen cannot be
moved by moving the
mouse.
The software is
malfunctioning.
Press Ctrl + Alt +
Delete on the keyboard
and exit from FLUOVIEW by
following the displayed
messages.
Then exit from windows and
reboot the computer.
1-2-1 & 1-2-11,
OPERATION
INSTRUCTIONS
19. The power supply of
Power Unit(FV5-PSU)
is not turned on.
The fuse has run out.
Contact your local Olympus
representative for assistance
then replace the huse.
-
20. The current limit
indicator on the UV Ar
laser power supply is
blinking.
The laser is in standby
condition.
Start the FLUOVIEW
software, then click the <On>
button for [UV Ar Laser] in
the [Laser Unit] group box in
the [Optical System
Configuration] window to
restart laser oscillation.
1-3-2-4,
OPERATION
INSTRUCTIONS
21. The error indicator on
the UV Ar laser power
supply will not turn off.
-
Consult your local Olympus
representative.
The battery maintains Bios
set value has run out.
Contact your local Olympus
representative for assistance
to replace the battery.
-
22. Bios some check error
message appears at the
moment of starting up
the computer.
Bios has broken.
Consult Olympus.
-
Summary of Contents for FLUOVIEW FV500
Page 2: ......
Page 8: ......
Page 10: ......
Page 24: ......
Page 110: ......
Page 327: ...APPLIED OPERATIONS Viewing 3D Image 2 217 Page Fig 2 106 Panel Showing Stereo 3D Images ...
Page 396: ......
Page 402: ......
Page 414: ......
Page 420: ......
Page 426: ......
Page 428: ......
Page 430: ......
Page 458: ......
Page 460: ......
Page 462: ......
Page 468: ......
Page 469: ......