2.2. Kr laser:
(1) Set the power switch to ON.
(2) Turn the key to the ON position. (The laser fan will start.)
(3) Turn the key further in the clockwise direction. (The key will return
to the ON position.) After the key is set to ON, it takes tens of
seconds before the laser oscillation starts.
2.3. HeNe (Green/Red) laser:
(1) Turn the key to the “I” (ON) position.
After the key is set to ON, it takes tens of seconds before the laser
oscillation starts.
To stabilize the laser beam output, we recommend leaving
the system for warm-up after turning the laser power ON.
The warm-up period should exceed 10 minutes when the
laser system in use uses an Ar laser or exceed 30 minutes
when it uses a Kr laser or Green HeNe laser.
2.4 UV-Ar laser
When using the system for the first time, check that the remote
cable is connected, the [POWER] switch on the front panel of the
chiller is set to “O-OFF”, its [COOL DOWN CYCLE] switch is set to
“I-ON” and the remote operation mode is set.
Fig. 1-3 Chiller Front Panel
From the second time and on, use the following method.
(1) Set [MAIN POWER] of the laser power supply to ON and turn the
key switch to the “I-ON” position.
(It takes about 30 seconds before the laser starts oscillation
together with a trigger sound. All error indicators turn off at this
time.)
HeNe (Green/Red) laser supply
UV laser power supply
NOTE
(1)
I-ON
O-OFF
Key switch
MAIN POWER
Output control
Power hour
counter
Error indicator
Summary of Contents for FLUOVIEW FV500
Page 2: ......
Page 8: ......
Page 10: ......
Page 24: ......
Page 110: ......
Page 327: ...APPLIED OPERATIONS Viewing 3D Image 2 217 Page Fig 2 106 Panel Showing Stereo 3D Images ...
Page 396: ......
Page 402: ......
Page 414: ......
Page 420: ......
Page 426: ......
Page 428: ......
Page 430: ......
Page 458: ......
Page 460: ......
Page 462: ......
Page 468: ......
Page 469: ......