Outline of LSM Observation Procedures
1-28
Page
1-2-3-3 Combination with IX
1. Turn the light path selector (1) on the right side of the microscope to
2. Engage the optimum cube for specimen dye by operating the cube
turret (2).
3. Focus on the specimen by looking into the eyepiece.
Be sure to adjust the diopter of the eyepiece in advance. (Refer to the
instruction manual of the IX50/70 microscope.)
When the Z motor is in use, clear the check mark in the
[Locked] check box in the [Z Stage] sub-panel in the
[Acquire] panel (see section 2-2-1-4-7 of this volume),
then focus on the specimen by operating the fine focus
adjustment knob of the microscope. Be sure not to
operate the fine focus adjustment knob while the
[Locked] check box is checked, for this may damage the
Z motor.
The specimen may float during oil-immersed
observation. In this case, prepare a stage clip (U-SCL)
and attach it to the microscope as shown on the left.
NOTE
(1)