Getting started FLUOVIEW/
Outline of LSM Observation Procedures
1-30
Page
5.
Look into the eyepiece and bring the specimen into focus.
Be sure to adjust the eyepiece diopter correctly (refer to the instruction manual for
the IX50/70).
When the Z-motor is in use, clear the check mark in the [Locked] check
box in the [Z Stage] sub-panel in the [Acquire] panel (see section 2-2-
1-4-7 of this volume) before bringing the specimen into focus using the
focus adjustment knobs of the microscope. If the fine focus
adjustment knob of the microscope is rotated while the [Locked] check
box is checked, the Z-motor may be destroyed.
The specimen may float during immersed observation. In this case,
attach the stage clip (U-SCL) to the microscope.
NOTE