Getting Started FLUOVIEW/
Outline of LSM Observation Procedures
1-37
Page
1-2-4-4 Combination with Upright Microscope (AX70A)
1. From the page tabs on the bottom right of the [Acquire] panel, select the [Optics]
sub-panel.
Fig. 1-9 [Optics] Sub-panel
2.
Select the <LSM> button in the [Light Path] group box.
The <LSM> button looks pushed in to indicate that it is selected.
(When scanning is started while the <BI> button is selected, the LSM light path is
selected automatically. It is switched back to the visual observation automatically
when scanning completes.)
3. When the FV5-ANA analyzer (1) is in use, disengage it from the light path by
setting the switch to the pulled-out position.
4. When only fluorescence observation is required, disengage the FV5-DICT
transmitted light DIC slider from the light path by pulling it out.
When the fluore transmitted light simultaneous interference observation is
required, the fluorescence resolution may be deteriorated slightly due to the
engagement of the FC5-DICT in the light path.
The cube turret is set automatically to
.
The optimum transmitted light DIC element for the objective in use is
engaged automatically in the light path.