FCC Information
FCC (U.S. Federal Communications Commission)
This equipment has been tested and found to comply with the limits for a Class B digital device,
pursuant to part 15 of the FCC Rules.
These limits are designed to provide reasonable protection against harmful interference in a
residential installation. This equipment generates, uses, and can radiate radio frequency energy,
and if not installed and used in accordance with the instructions, may cause harmful interference
to radio communications.
However, there is no guarantee that interference will not occur in a particular installation. If this
equipment does cause unacceptable interference to radio or television reception, which can
be determined by turning the equipment off and on, the user is encouraged to try to correct the
interference by one or more of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and receiver.
• Connect the equipment into an outlet on a circuit different from that to which the receiver is
connected.
• Consult your re-seller or an experienced radio/TV technician for help.
FCC Warning:
To assure continued FCC compliance, the user must use a grounded power supply cord and the
provided shielded video interface cable with bonded ferrite cores. Also, any unauthorized changes
or modifications to this monitor would void the user’s authority to operate this device.
Test of medical power adapter
The power supply is Class I. Protective earthed parts are only located inside the power supply
enclosure and therefore are not accessible. The accessible parts of the enclosure and output
are DOUBLE INSULATED from all mains voltage electrical parts. Ground continuality test is not
relevant.
Leakage test of the adapter is made by having the measurement equipment connected to the AC
plug, no need to open the plastic enclosure.