Page 1-10
OLYMPIC AUTO-LENGTH
™
Installation
Canceling Calibration Mode
If the Auto-Length is inadvertently placed in calibration mode,
cancel the mode by pressing OFF at any time
before
taking the
calibration measurement.
Testing for Correct Operation
Before placing the Auto-Length in service, tested it as follows:
1. Place an object in the tray so that it is touching the end of the tray
furthest from the measuring bar.
2. Press MEASURE.
3. After a two second delay, the Auto-Length will take three measure-
ments. The measurements should all be within ±1 mm.
4. If the measurements are not within ±1 mm, recalibrate the Auto-
Length and remeasure. If the measurements are beyond speci-
fications, see
Troubleshooting
on page 3-18.
Electrical Requirements
CAUTION: This device must be connected directly to a
hospital-grade electrical outlet.
The Auto-Length is rated at 120 Vac, 60 Hz, 0.4 amperes, and must
be connected to a hospital-grade electrical outlet.
Battery Operation
The device can be powered by an internal rechargable battery pack
consisting of nine sub-c NiCad batteries rated at 10.8 Volts (1.2 Volts
per cell).
For information on charging and replacing batteries, see
Replacing
the Battery Pack
on page 3-1.