4-3
Inspecting the Bili-Bassinet
4
Clean the phototherapy pad compartment:
a
Using a cloth dampened with disinfectant, wipe the compartment. Be sure to
clean beneath wires and counterbalance springs.
b
Sparingly wipe the two exposed counterbalance shafts on either side of the
compartment with petroleum jelly or Dow Corning
®
111 (silicone).
5
Place the phototherapy pad fixture:
a
Reattach the two connectors to the bottom of the phototherapy pad, then
twist their rings to lock them in place.
b
Seat the phototherapy pad fixture into the bassinet. Ensure that the fixture is
securely seated.
c
Place the mattress. Ensure that the wire handles are fully seated on either
side of the fixture.
6
Test the phototherapy pad for proper operation (see page 2-3).
Inspecting the Bili-Bassinet
To inspect the Bili-Bassinet:
Examine both the arm pivot assemblies and the counterbalance mechanisms of
the overhead light fixtures for any wear and looseness and to verify smooth
operation.
Test the transparent sides for correct operation: move each transparent side into
the down position, then confirm that each side locks in the up position.
Replacing the Bulbs
Replace the Bili-Bassinet bulbs at 3,000 hours to maintain greater than 75% of
initial irradiance (irradiance measured at 100 hours).
Bulb Timers
The bulb timers operate whenever the bulbs illuminate. View the elapsed time
shown to track bulb life and determine when to replace the treatment bulbs. The
timers for each of the overhead lights and phototherapy pad are located to the right
of the mattress tilt lever.
CAUTION
Perform the inspection annually to assure patient and operator safety.
NOTE
#
The
elapsed time
bulb timers cannot be reset by the user.
The bulb timers may be labeled as
Lamp Timer
.
Change all treatment bulbs at the same time.
Record the bulb change in the
Medical Device History
.
CAUTION
Use only bulbs specified by Olympic Medical. Use of other bulbs may affect the safety and
efficacy of treatment.