6-4
OLYMPIC BILI-BASSINET
Instruction Manual
Repair Procedures
6
Reassemble the Bili-Bassinet:
a
Insert the ballast box cover. Ensure that the tabs on the cover align with
the ballast box slots (see Figure 6.3). Secure the ballast box cover with its
two screws.
b
Attach the eight connectors to the ballast box.
c
Place the panel, then secure in place with its four screws.
7
Test the device for operation (see page 3-2).
Replacing the Switches
If the switches malfunction, they may need to be replaced.
Required items:
Screwdriver
To replace the switches:
1
Press the power switches to the
OFF
(O) position, then unplug the power cord.
2
Lower the overhead lights.
3
Remove the phototherapy pad fixture (see Figure 6.4):
a
Remove the mattress, then pull the wire handles up from their recesses in
the phototherapy pad fixture.
b
Pull the
MATTRESS TILT
handle while lifting the phototherapy pad fixture
by the wire handle at the front of the fixture. Next, release the
MATTRESS
TILT
handle, then grasp the wire handle at the back of the fixture and lift the
fixture clear of the bassinet.
c
Twist the two connector rings to remove them from the bottom of the
phototherapy pad fixture.
NOTE
#
The cover fits outside of the box sides. Removing the rear cover will make it easier to locate
the tabs in the slots but is not strictly necessary.