31
Use
– Power 200 W
– Effect 2
– Mode: <SalineCoag>
– Power 120 W
– Effect 2
●
Check that the correct setting <Saline> appears on the
display.
●
If the correct setting is not displayed, disconnect the plug and
try again.
●
Do not pull on the cable to disconnect the plug. Always pull
on the plug to avoid arcing.
The presettings may not produce optimum results for all tissue
types and procedures. If necassary, adjust the settings to the
procedure and the required tissue effect.
Duration of activation
Do not exceed the maximum time of activation. For the specific
maximum activation time, refer to the instructions for use of the
compatible instruments.