Cleaning and disinfection
If other detergents are used, select immersion times,
concentrations and temperatures of the solutions as instructed by
the manufacturer of the detergent.
8.2.3 Manually cleaning the product
A complete manual cleaning cycle consists of 3 steps:
– Cleaning
– Rinsing
– Drying
Specific instructions for these steps are given on the following
pages.
Cleaning
1. Create a detergent solution as instructed by the manufacturer
of the detergent.
2. Completely immerse the HF cable in the detergent solution
for 20 minutes.
Make sure that there are no air bubbles on the HF cable
while immersed.
3. Thoroughly brush every gap at least 10 times with a surface
brush.
4. Using a syringe, thoroughly rinse the HF cable and every gap
at least 10 times with 10 ml of detergent solution.
Rinsing
1. Select a rinsing device (e.g. a cleaning pistol).
The rinsing device must be suitable for cleaning medical
devices and deliver a minimum pressure of 1 bar (14.5 psi).
2. Connect the rinsing device to a supply of deionized water. Do
not use tap water for rinsing because it can be chlorinated.
3. Thoroughly rinse the HF cable with deionized water for at
least 30 seconds using the rinsing device.
Drying
1. Let the HF cable dry at room temperature.
To support drying, use a clean lint-free cloth or filtered
compressed air.