Cleaning and disinfection
Specific instructions for these steps are given on the following
pages.
Disinfection
1. Create a disinfectant solution as instructed by the
manufacturer of the disinfectant.
2. Completely immerse the HF cable in the disinfectant for
45 minutes.
Make sure that there are no air bubbles on the products while
immersed.
3. Remove the HF cable from the disinfectant.
Rinsing
1. Completely immerse the HF cable in 8 liters of sterile
deionized water for at least 1 minute.
2. Rinse all lumens with at least 100 ml of sterile deionized
water.
3. Repeat steps 1 and 2 twice for a total of 3 rinses. Always use
fresh portions of sterile deionized water for each rinse.
Drying
1.
Dry the product using a clean lint-free cloth or filtered
compressed air for at least 10 seconds until the HF cable is
visually dry.
2. Check that no residual rinse water remains on the products.
8.4 Automated cleaning and thermal disinfection
Automated cleaning and thermal disinfection removes patient
debris from the product. This is a prerequisite for the subsequent
sterilization steps.
Use only washer-disinfectors with thermal disinfection programs
in accordance with ISO 15883 for automated cleaning and
thermal disinfection of the product.