Teardown, Cleaning, and Reassembly of the Miscellaneous Parts of the Olympus BH-2 Microscope Frames
Revision 1
Page 19 of 45
Figure 56
–
Thoroughly clean slide-mount base (old style)
The older style nosepiece slide mount is now ready to
be reinstalled onto the stand and put back into service
(see
Figure 57
–
Slide-mount parts ready for use (old style)
Servicing the New-Style Slide Mount (BHS/BHT)
This section applies only to the newer style assembly
(see
). The older style assembly is covered in
the previous section.
Disassemble the New-Style Slide Mount (BHS/BHT)
Using a suitable slotted screwdriver, loosen and remove
the stop screw for the nosepiece slide-mount locking
block (see
Figure 58
–
Remove the locking-block stop screw
Using a suitable slotted screwdriver, loosen and remove
the pivot screw for the nosepiece slide-mount locking
block (see
Figure 59
–
Remove the locking-block pivot screw
Separate the locking block from the slide-mount base
(see
Figure 60
–
Separate locking block from slide-mount base
Clean Components of New-Style Slide Mount (BHS/BHT)
Using a cotton swab and a suitable solvent (acetone),
thoroughly clean the nosepiece slide-mount base,
paying special attention to the surface upon which the
locking block slides and the dovetail bearing surfaces.
(see
Figure 61
–
Thoroughly clean slide-mount base (new style)