Teardown, Cleaning, and Reassembly of the Miscellaneous Parts of the Olympus BH-2 Microscope Frames
Revision 1
Page 21 of 45
Reinstall the Nosepiece Slide Mount (BHS/BHT)
Reinstall the nosepiece slide mount onto the BHS/BHT
pillar arm per the following.
Reinstall the Slide-Mount Block
Place the nosepiece slide mount into position on the
BHS/BHT pillar arm. See
Figure 68
–
Place the nosepiece slide mount into position
Carefully align the three holes in the viewing head
recess of the BHS/BHT pillar arm with the three tapped
holes in the nosepiece slide mount base. See
Figure 69
–
Align drilled holes with tapped holes
Using a suitable JIS screwdriver, reinstall the three
m2.6x5 screws to secure the nosepiece slide mount to
the BHS/BHT pillar arm (see
Figure 70
–
Secure nosepiece slide mount with screws
Reinstall the Slide-Mount Locking Thumbscrew
Apply a light coating of grease to the threads of the
nosepiece slide-mount locking thumbscrew (see
). Wipe away any excess grease.
Figure 71
–
Lightly grease the thumbscrew threads
Carefully reinstall the nosepiece locking thumbscrew
into the tapped hole in the BHS/BHT pillar arm (see
). Wipe away any visible grease squeeze-out.
Figure 72
–
Reinstall the nosepiece locking thumbscrew
Nosepiece Slide Mount is Ready for Use
The nosepiece slide mount and lock screw are now
ready to provide years of trouble-free service (see
Figure 73
–
Newly serviced slide mount ready for use