Teardown, Cleaning, and Reassembly of the Miscellaneous Parts of the Olympus BH-2 Microscope Frames
Revision 1
Page 37 of 45
Servicing the Light Exit Assembly
The Light Exit Assembly can be serviced per the
following.
Disassemble the Light Exit Assembly
Rotate the knurled field-adjustment ring to the fully
clockwise position to set the field diaphragm to the
wide-open position (see
Figure 155
–
Fully open the field diaphragm
Using a suitable JIS screwdriver, remove the three m2x7
screws securing the filter holder in place in the recess of
the knurled field-adjustment ring (see
Figure 156
–
Remove screws securing the filter holder
Lift and remove the filter holder from the recess in the
knurled field-adjustment ring (see
Figure 157
–
Lift and remove the filter holder
Lift and remove the knurled field-adjustment ring from
the frame of the Light Exit Assembly (see
Figure 158
–
Lift and remove the field-adjustment ring
Use a suitable lens spanner tool (item 8 of
to loosen the threaded retaining ring holding the glass
field-exit lens in place (see
and
Figure 159
–
Lens spanner tool with bent tips
Figure 160
–
Loosen the threaded retaining ring