DMTA-10045-01EN, Rev. E, August 2016
Applications
123
Figure 6
‑
1 Materials — skin
‑
to
‑
core disbonds in flat or constant geometry
The following products are used in this procedure:
•
Honeycomb composite training standard: 25 mm (1 in.) thick, 6-ply CFRP top
skin and 3-ply fiberglass lower skin. Includes a 25 mm (1 in.) disbond on each
side. P/N: NEC-6407 [U8862302]
•
Cable for pitch-catch and MIA modes, 1.83 m (6 ft).
P/N: SBM-CPM-P11 [U8800058]
•
Pitch-catch probe for general use; tip spacing 14 mm. P/N: S-PC-P14 [U8800601]
To set the initial BondMaster 600 configuration
1.
Connect the probe and the cable to the PROBE connector of the BondMaster 600.
2.
When prompted, press
CONTINUE
(A key) to accept the PowerLink probe
information.
If you are using a non-PowerLink probe, access the
APPL SELECT
menu (A key)
using the ADV SETUP menu key (
).
3.
Select the
Skin to Core Disbonds (Flat)
application, and then press
to accept
NOTE
Summary of Contents for BondMaster 600
Page 8: ...DMTA 10045 01EN Rev E August 2016 Table of Contents viii ...
Page 10: ...DMTA 10045 01EN Rev E August 2016 List of Abbreviations x ...
Page 26: ...DMTA 10045 01EN Rev E August 2016 Important Information Please Read Before Use 16 ...
Page 28: ...DMTA 10045 01EN Rev E August 2016 Introduction 18 ...
Page 32: ...DMTA 10045 01EN Rev E August 2016 Chapter 1 22 ...
Page 72: ...DMTA 10045 01EN Rev E August 2016 Chapter 3 62 ...
Page 78: ...DMTA 10045 01EN Rev E August 2016 Chapter 4 68 Figure 4 2 Crosshairs and the NULL point ...
Page 130: ...DMTA 10045 01EN Rev E August 2016 Chapter 5 120 ...
Page 216: ...DMTA 10045 01EN Rev E August 2016 Chapter 7 206 ...
Page 230: ...DMTA 10045 01EN Rev E August 2016 Appendix A 220 ...
Page 234: ...DMTA 10045 01EN Rev E August 2016 Appendix B 224 ...
Page 240: ...DMTA 10045 01EN Rev E August 2016 List of Figures 230 ...
Page 242: ...DMTA 10045 01EN Rev E August 2016 List of Tables 232 ...