DMTA-10045-01EN, Rev. E, August 2016
Chapter 6
124
Figure 6
‑
2 The Skin To Core Disbonds (Flat) application
To calibrate the signals
1.
Press the MAIN menu key (
) once, and set the
FREQ
(A key) to
11 kHz
using the knob.
2.
Set the probe tips over a good area of the standard, adjust the gain using the
GAIN key (
), and then ensure that the signal in the RF view (left) is contained
between 1 to 2 vertical divisions (see Figure 6-3 on page 124).
Figure 6
‑
3 The GAIN adjustment to obtain a contained signal