DMTA-10045-01EN, Rev. E, August 2016
Applications
143
The following products are used in this procedure:
•
Honeycomb composite training standard: 25 mm (1 in.) thick, 3 and 6 ply CFRP
top skin, and 3 ply fiberglass bottom skin. Includes one 13 mm (0.5 in.) and one
25 mm (1 in.) disbond on each side. P/N: NEC-6433 [U8620490].
•
Cable for pitch-catch and MIA modes, 1.83 m (6 ft).
P/N: SBM-CPM-P11 [U8800058]
•
Right angle MIA probe, 13 mm (0.5 in.) tip. P/N: S-MP-3 [U8010011]
To set the initial BondMaster 600 configuration
1.
Connect the probe and the cable to the PROBE connector of the BondMaster 600.
2.
When prompted, press
CONTINUE
(A key) to accept the PowerLink probe
information.
If you are using a non-PowerLink probe, access the
APPL SELECT
menu (A key)
using the ADV SETUP menu key (
).
3.
Select the
Smaller Disbonds and Repair Identification
application, and then
press
to accept (see Figure 6-30 on page 143).
Figure 6
‑
30 The Smaller Disbonds and Repair Identification application
NOTE
Summary of Contents for BondMaster 600
Page 8: ...DMTA 10045 01EN Rev E August 2016 Table of Contents viii ...
Page 10: ...DMTA 10045 01EN Rev E August 2016 List of Abbreviations x ...
Page 26: ...DMTA 10045 01EN Rev E August 2016 Important Information Please Read Before Use 16 ...
Page 28: ...DMTA 10045 01EN Rev E August 2016 Introduction 18 ...
Page 32: ...DMTA 10045 01EN Rev E August 2016 Chapter 1 22 ...
Page 72: ...DMTA 10045 01EN Rev E August 2016 Chapter 3 62 ...
Page 78: ...DMTA 10045 01EN Rev E August 2016 Chapter 4 68 Figure 4 2 Crosshairs and the NULL point ...
Page 130: ...DMTA 10045 01EN Rev E August 2016 Chapter 5 120 ...
Page 216: ...DMTA 10045 01EN Rev E August 2016 Chapter 7 206 ...
Page 230: ...DMTA 10045 01EN Rev E August 2016 Appendix A 220 ...
Page 234: ...DMTA 10045 01EN Rev E August 2016 Appendix B 224 ...
Page 240: ...DMTA 10045 01EN Rev E August 2016 List of Figures 230 ...
Page 242: ...DMTA 10045 01EN Rev E August 2016 List of Tables 232 ...