DMTA-10045-01EN, Rev. E, August 2016
Chapter 6
144
To calibrate the signals
1.
Press the DISP/DOTS menu key (
) once, and then set the
POSITION
(C key)
to
CENTER
.
2.
Press the ALARM menu key (
), and then set the
XY ALM 1
(B key) to
OFF
.
3.
Place the probe tip on a good part of the standard, and then press the CAL NULL
key (
).
4.
Lift the probe into the air to observe the dot’s movement; if the dot disappears off
the screen, press the GAIN key (
), and then use the knob to adjust the dot
position so that it remains on the screen (see Figure 6-31 on page 144).
Figure 6
‑
31 Adjusting the dot position
5.
Place the probe tip on a good part of the standard, and then press the CAL NULL
key (
).
6.
Slowly scan over a disbond and a repaired area, and then press the FREEZE key
(
) [see Figure 6-32 on page 145].
Summary of Contents for BondMaster 600
Page 8: ...DMTA 10045 01EN Rev E August 2016 Table of Contents viii ...
Page 10: ...DMTA 10045 01EN Rev E August 2016 List of Abbreviations x ...
Page 26: ...DMTA 10045 01EN Rev E August 2016 Important Information Please Read Before Use 16 ...
Page 28: ...DMTA 10045 01EN Rev E August 2016 Introduction 18 ...
Page 32: ...DMTA 10045 01EN Rev E August 2016 Chapter 1 22 ...
Page 72: ...DMTA 10045 01EN Rev E August 2016 Chapter 3 62 ...
Page 78: ...DMTA 10045 01EN Rev E August 2016 Chapter 4 68 Figure 4 2 Crosshairs and the NULL point ...
Page 130: ...DMTA 10045 01EN Rev E August 2016 Chapter 5 120 ...
Page 216: ...DMTA 10045 01EN Rev E August 2016 Chapter 7 206 ...
Page 230: ...DMTA 10045 01EN Rev E August 2016 Appendix A 220 ...
Page 234: ...DMTA 10045 01EN Rev E August 2016 Appendix B 224 ...
Page 240: ...DMTA 10045 01EN Rev E August 2016 List of Figures 230 ...
Page 242: ...DMTA 10045 01EN Rev E August 2016 List of Tables 232 ...