DMTA-10045-01EN, Rev. E, August 2016
Chapter 6
150
To calibrate the signals
1.
Install a backing mat or foam under the standard. Doing so will make the
readings more stable.
2.
Pour a generous amount of couplant onto the standard.
3.
Place the probe on a good area of the standard, and then press the CAL NULL key
(
).
4.
Slowly glide the probe over the first disbond and hold it there.
5.
Press the DISP/DOTS menu key (
) twice to display the DOTS screen.
6.
Press
STORE NEXT
(A key) to record the first dot (see Figure 6-40 on page 150).
Figure 6
‑
40 The first dot recorded
7.
Slowly glide the probe over the second disbond, and then press
STORE NEXT
(A key) to record the second dot (see Figure 6-41 on page 151).
Summary of Contents for BondMaster 600
Page 8: ...DMTA 10045 01EN Rev E August 2016 Table of Contents viii ...
Page 10: ...DMTA 10045 01EN Rev E August 2016 List of Abbreviations x ...
Page 26: ...DMTA 10045 01EN Rev E August 2016 Important Information Please Read Before Use 16 ...
Page 28: ...DMTA 10045 01EN Rev E August 2016 Introduction 18 ...
Page 32: ...DMTA 10045 01EN Rev E August 2016 Chapter 1 22 ...
Page 72: ...DMTA 10045 01EN Rev E August 2016 Chapter 3 62 ...
Page 78: ...DMTA 10045 01EN Rev E August 2016 Chapter 4 68 Figure 4 2 Crosshairs and the NULL point ...
Page 130: ...DMTA 10045 01EN Rev E August 2016 Chapter 5 120 ...
Page 216: ...DMTA 10045 01EN Rev E August 2016 Chapter 7 206 ...
Page 230: ...DMTA 10045 01EN Rev E August 2016 Appendix A 220 ...
Page 234: ...DMTA 10045 01EN Rev E August 2016 Appendix B 224 ...
Page 240: ...DMTA 10045 01EN Rev E August 2016 List of Figures 230 ...
Page 242: ...DMTA 10045 01EN Rev E August 2016 List of Tables 232 ...