DMTA-10045-01EN, Rev. E, August 2016
Chapter 6
162
3.
Select the
Skin To Core Disbonds (Tapered)
application, and then press
to
accept (see Figure 6-56 on page 162).
Figure 6
‑
56 The Skin To Core Disbonds (Tapered) application
4.
Press the MAIN menu key (
) once, and set the
SW. RATE
(E key) to
LOW
.
When developing an application or procedure, the lower sweep rate setting
usually provides the best results.
5.
If necessary, adjust the
START FREQ
(C key) and
STOP FREQ
(D key) values.
As a general starting point, values in the range of 5 kHz to 50 kHz are usually
very suitable.
6.
Press the RUN key (
) once to display the
SPEC+XY
RUN.
7.
Press the MAIN menu key (
) again to display the
SPECIAL
screen.
Summary of Contents for BondMaster 600
Page 8: ...DMTA 10045 01EN Rev E August 2016 Table of Contents viii ...
Page 10: ...DMTA 10045 01EN Rev E August 2016 List of Abbreviations x ...
Page 26: ...DMTA 10045 01EN Rev E August 2016 Important Information Please Read Before Use 16 ...
Page 28: ...DMTA 10045 01EN Rev E August 2016 Introduction 18 ...
Page 32: ...DMTA 10045 01EN Rev E August 2016 Chapter 1 22 ...
Page 72: ...DMTA 10045 01EN Rev E August 2016 Chapter 3 62 ...
Page 78: ...DMTA 10045 01EN Rev E August 2016 Chapter 4 68 Figure 4 2 Crosshairs and the NULL point ...
Page 130: ...DMTA 10045 01EN Rev E August 2016 Chapter 5 120 ...
Page 216: ...DMTA 10045 01EN Rev E August 2016 Chapter 7 206 ...
Page 230: ...DMTA 10045 01EN Rev E August 2016 Appendix A 220 ...
Page 234: ...DMTA 10045 01EN Rev E August 2016 Appendix B 224 ...
Page 240: ...DMTA 10045 01EN Rev E August 2016 List of Figures 230 ...
Page 242: ...DMTA 10045 01EN Rev E August 2016 List of Tables 232 ...