DMTA-10045-01EN, Rev. E, August 2016
Chapter 6
170
a)
Prioritize negative peaks over positive peaks.
b)
In case you have multiple peaks, prioritize the first (left) peak over the peaks
on the right side.
c)
When in doubt, repeat the frequency selection steps and ensure that you
maintain a constant pressure on the probe. If your results were unsatisfactory,
this might have been because you did not hold the probe properly.
d)
The use of a probe holder is NOT recommended because it can impact the
mechanical impedance of the area being scanned.
Figure 6
‑
65 Selecting the best operating frequency
7.
When you have obtained a satisfactory frequency, press
DONE
(E key).
To calibrate the signals
1.
With the frequency properly selected, place the probe on a defect-free portion of
the standard and then press the CAL NULL key (
).
NOTE
Summary of Contents for BondMaster 600
Page 8: ...DMTA 10045 01EN Rev E August 2016 Table of Contents viii ...
Page 10: ...DMTA 10045 01EN Rev E August 2016 List of Abbreviations x ...
Page 26: ...DMTA 10045 01EN Rev E August 2016 Important Information Please Read Before Use 16 ...
Page 28: ...DMTA 10045 01EN Rev E August 2016 Introduction 18 ...
Page 32: ...DMTA 10045 01EN Rev E August 2016 Chapter 1 22 ...
Page 72: ...DMTA 10045 01EN Rev E August 2016 Chapter 3 62 ...
Page 78: ...DMTA 10045 01EN Rev E August 2016 Chapter 4 68 Figure 4 2 Crosshairs and the NULL point ...
Page 130: ...DMTA 10045 01EN Rev E August 2016 Chapter 5 120 ...
Page 216: ...DMTA 10045 01EN Rev E August 2016 Chapter 7 206 ...
Page 230: ...DMTA 10045 01EN Rev E August 2016 Appendix A 220 ...
Page 234: ...DMTA 10045 01EN Rev E August 2016 Appendix B 224 ...
Page 240: ...DMTA 10045 01EN Rev E August 2016 List of Figures 230 ...
Page 242: ...DMTA 10045 01EN Rev E August 2016 List of Tables 232 ...