DMTA-10045-01EN, Rev. E, August 2016
Chapter 6
132
Figure 6
‑
15 RUN 5 — SCAN
To obtain clearer phase readings in the SCAN views, it is recommended to NULL the
probe in the air.
To fine-tune the instrument settings
1.
Depending on your requirements, set the alarm box parameters, horn, or external
horn (louder). For more details about alarms, see “Alarms, Connectivity, and
Memory Specifications” on page 217.
2.
Depending on your requirements, change your real-time readings.
The default real-time readings display the XY signal’s live amplitude and phase.
For information on how to change the real-time readings, see “Displaying Real-
Time Readings” on page 58.
The list of all parameters is shown in Figure 6-16 on page 133.
TIP
Summary of Contents for BondMaster 600
Page 8: ...DMTA 10045 01EN Rev E August 2016 Table of Contents viii ...
Page 10: ...DMTA 10045 01EN Rev E August 2016 List of Abbreviations x ...
Page 26: ...DMTA 10045 01EN Rev E August 2016 Important Information Please Read Before Use 16 ...
Page 28: ...DMTA 10045 01EN Rev E August 2016 Introduction 18 ...
Page 32: ...DMTA 10045 01EN Rev E August 2016 Chapter 1 22 ...
Page 72: ...DMTA 10045 01EN Rev E August 2016 Chapter 3 62 ...
Page 78: ...DMTA 10045 01EN Rev E August 2016 Chapter 4 68 Figure 4 2 Crosshairs and the NULL point ...
Page 130: ...DMTA 10045 01EN Rev E August 2016 Chapter 5 120 ...
Page 216: ...DMTA 10045 01EN Rev E August 2016 Chapter 7 206 ...
Page 230: ...DMTA 10045 01EN Rev E August 2016 Appendix A 220 ...
Page 234: ...DMTA 10045 01EN Rev E August 2016 Appendix B 224 ...
Page 240: ...DMTA 10045 01EN Rev E August 2016 List of Figures 230 ...
Page 242: ...DMTA 10045 01EN Rev E August 2016 List of Tables 232 ...