DMTA-10045-01EN, Rev. E, August 2016
Specifications
215
Table 13 Pitch
‑
catch tone burst and swept mode specifications
Category
Parameter
Value
Pitch-catch tone burst
Display modes (RUN
key)
Choice of RF view (time-based
waveform featuring raw waveform
named RF, or signal amplitude
envelope named IMPULSE), single
impedance plane (also named XY
flying dot), split screen (RF and XY),
strip chart (named SCAN, representing
amplitude and phase against time),
split screen (XY + SCAN)
Frequency range
1 kHz to 50 kHz
Gain
RF waveform gain (raw signal): 0 dB to
70 dB, adjustable in 0.1 dB or 1 dB
increments. An additional 0 dB to
60 dB is available for the XY flying dot
display.
Width
360
μ
s
to 10 ms, adjustable in 50
μ
s
increments
Gate
10
μ
s to 7920
μ
s, adjustable in 10
μ
s
steps. New AUTO gate mode that
auto-detects the peak amplitude from
RF signal.
Cycles
1 to 10, adjustable in 1 cycle increments
Repetition rate
5 to 500 repetitions per second,
adjustable in 5 rep/s increments
Dot recordings
Up to 25 user-defined dot recordings
Summary of Contents for BondMaster 600
Page 8: ...DMTA 10045 01EN Rev E August 2016 Table of Contents viii ...
Page 10: ...DMTA 10045 01EN Rev E August 2016 List of Abbreviations x ...
Page 26: ...DMTA 10045 01EN Rev E August 2016 Important Information Please Read Before Use 16 ...
Page 28: ...DMTA 10045 01EN Rev E August 2016 Introduction 18 ...
Page 32: ...DMTA 10045 01EN Rev E August 2016 Chapter 1 22 ...
Page 72: ...DMTA 10045 01EN Rev E August 2016 Chapter 3 62 ...
Page 78: ...DMTA 10045 01EN Rev E August 2016 Chapter 4 68 Figure 4 2 Crosshairs and the NULL point ...
Page 130: ...DMTA 10045 01EN Rev E August 2016 Chapter 5 120 ...
Page 216: ...DMTA 10045 01EN Rev E August 2016 Chapter 7 206 ...
Page 230: ...DMTA 10045 01EN Rev E August 2016 Appendix A 220 ...
Page 234: ...DMTA 10045 01EN Rev E August 2016 Appendix B 224 ...
Page 240: ...DMTA 10045 01EN Rev E August 2016 List of Figures 230 ...
Page 242: ...DMTA 10045 01EN Rev E August 2016 List of Tables 232 ...