DETAILS OF EACH OPERATION
/ Camera focus alignment
II.
DSU
II .
3 - 1 5
Page
3-14 Camera focus alignment
1. Bring a live image on monitor. Put objective lens 10x into light path
and displace the specimen from light path.
2. Press left IN button (1) of DSU hand switch and put the disk into light
path.
3.
Stop the disk turning.
1) Switch off power of control box.
2) Pull out disk control cable (2) from disk box.
3) Switch on power of control box.
*
Upon completion of camera focus alignment, reverse the
procedures to go back to the original state.
4. Turn rear side turret and bring arbitrarily selected fluorescent mirror
unit or empty place into light path.
5.
Turn front side turret and bring reference mirror of No.1 into light path.
6. Adjust transmitting light for enabling to see the strips of disk on
monitor.
7.
Loosen fixing screw (3) located, using the Allen screwdriver.
8. Seeing live image on monitor, turn adjusting ring of mount adapter (4)
and bring strips of disk in focus.
9. Tighten the fixing screw (3) (at least one place) located at port, using
the Allen screwdriver.
3-15 Power shutdown
1.
Switch off the lamp.
2.
Switch off the control box.
3.
Switch off the CCD camera.
(1)
(3)
(2)
(4)