28
BX53/51/41-P
6. Condenser (U-POC-2)
Aplanatic/achromatic condenser, top lens swing-out type
Polarizer 360° rotatable
0° position adjustable
N. A . 0.9 (top lens swung in)
Applicable objectives: 2X to 100X (2X to 4X objectives are used with the top lens
swung out)
Operating environment
· Indoor use
· Altitude up to 2000 m
· Temperature: 5° to 40°C (41° to 104°F)
· Maximum relative humidity 80% for temperatures up to 31°C (88°F)
decreasing linearly to 50% relative humidity at 40°C (104°F)
· Main supply voltage fluctuations not to exceed ±10% of the nominal voltage
· Installation/Overvoltage Category II (In accordance with IEC60664)
· Pollution Degree 2 (In accordance with lEC60664)
5. Stage (U-SRG2)
Type: Rotatable stage with 3-point centering mechanism 360° horizontal rotation,
clamps at any desired position
360° scale (minimum division: 1° reading)
· Stage clips (U-SCB2) attachable for holding specimen in place
· Attachable mechanical stage (U-FMP)
Item
Specifications
Intermediate attachment conoscopic and
orthoscopic observation (U-CPA)
Intermediate attachment for
orthoscopic observation (U-OPA)