110
5
Attaching the mercury burner
Applicable burner USH-103OL (made by Ushio Inc.)
Loosen the clamping screw on the top of the lamp housing using
the Allen screwdriver.
1
Fit the halogen lamp housing from the above and while pressing down
the fixing screw, tighten it with the Allen screwdriver.
Cautions when replacing the bulb during observation:
The bulb, the lamp housing and areas around the lamp
housing are extremely hot during and right after use.
Set the main switch to (OFF) and disconnect the power
cord. Wait until the bulb, the lamp housing and areas
around the lamp housing are sufficiently cooled down and
replace the old bulb with the applicable bulb.
5
CAUTION
2
4
c
b
a
b
1
Hold the upper section of the lamp housing and pull it upward to
remove the socket part.
To prevent malfunctions, do not hold the lamp housing by
the centering knobs
a
.
2
NOTE
Place the socket part so that the radiation fin b faces down.
Loosen the lamp clamping screws (2 pcs.) of the socket part.
Hold the mercury burner c attached and remove the lower side from
the mount first and then remove the upper side.
The socket part is attached with either one of followings:
· Dummy burner for transportation (factory default)
· Old burner (to be replaced)
3
4
5
TIP