82
: Turns ON. : Blinks. :Turns OFF. : Varies depending on state.
Phenomena
Cause
Remedy
Page
Number of beeps
CB
LIM /
HS
MIX
None
/
(Either one of them turns ON.)
–
The cable of the coded nosepiece is not connected.
Set the main switch to (OFF) and connect the cable of the
coded nosepiece.
16, 115
h) The system cannot be controlled by PC.
None
/
–
–
The interface cable is not connected.
Connect the interface cable.
121
The main switch is set to (OFF) while the
communication between PC and the microscope is in
process.
Close the application software.
Then, set the main switch to (ON) and start the application
software.
16, –
The device driver is not installed.
Install the device driver from the installation media of the
application software.
–
The device is not detected properly.
Scan the device again with the Windows
®
device manager.
–
The USB cable is connected to the USB connector of PC
for the first time.
Install the device driver in accordance with the installation
wizard.
–
The communication between the control box (BX3M-CB)
and PC is disconnected.
Remove the interface cable from PC and connect it again. Then,
restart the application software.
If the communication is still not restored, turn OFF the power
supplies of the microscope and PC and turn ON again.
16, 121
i) The beep sound is heard once suddenly, and the system returns to the state that the power is ON.
1 time
/
–
–
The large supply voltage fluctuation, etc. occurs caused by
thunder, etc. and the system was reset.
This is not a failure.
If this phenomenon occurs frequently, contact Olympus.
–
j) The beep sound is heard five times.
5 times
/
–
–
The communication between the control box (BX3M-CB) and
the hand switch (BX3M-HS) is disconnected.
The cable of the hand switch (BX3M-HS) is removed while
the power is ON.
Set the main switch to (OFF) once and set the main switch to
(ON) again.
If the error occurs frequently, contact Olympus.
16
5 times
/
–
–
The communication between the control box (BX3M-CB)
and the hand switch for motorized nosepiece (BX3M-HSRE)
is disconnected.
The cable of the hand switch for motorized nosepiece (BX3M-
HSRE) is removed while the power is ON.
Set the main switch to (OFF) once and set the main switch to
(ON) again.
If the error occurs frequently, contact Olympus.
16
5 times
/
–
–
The communication between the control box (BX3M-CB) and
the hand switch for exposure (U-HSEXP) is disconnected.
The cable of the hand switch for exposure (U-HSEXP) is
removed while the power is ON.
Set the main switch to (OFF) once and set the main switch to
(ON) again.
If the error occurs frequently, contact Olympus.
16
5 times
/
–
–
The communication between the control box (BX3M-CB)
and PC is disconnected.
Set the main switch to (OFF) once and set the main switch to
(ON) again.
If the error occurs frequently, contact Olympus.
16