Safety Check Instruction

 

 

 

 

 

 

 

©  Celon AG 

Page 1 / 9 

 

Issue date: 01.09.2008 

SCI.991007-1.4 

Safety Check 

Instruction.doc 

Product name: 

Hardware version: 

CelonLab 

ENT 

(100...120 V~) 

 30

 

REF no.:  WB991007 

LAB991.025.007 

The unit must undergo a safety check at annual intervals in accordance with the national statutory regulations. Please follow these test 
instructions. All tests must be done with fully functional and calibrated test equipment and by technicians trained in the service / maintenance 
of electrical medical devices. 

Record the test results in the “safety check report” for reference in future tests and provide the user of the unit 

with a signed report. If the unit fails to meet any of the checks, contact the manufacturer. 

During service / maintenance take care of the different hardware versions which may be applicable (see below). 

 

The 

hardware version

 of the unit can be identified by the serial number which can be found on the type plate at the back side of the unit. The 

table below shows the serial number range to identify the hardware version. 

Hardware version 

Serial number (SN) 

30 … 31 

A22701LE01.0001 ... A58702LE01.9999 

32 … ≥ 35 

1000-

1001 … 6172-9999 

From hardware version 35 the serial number itself includes already the hardware version. This code was introduced for units manufactured 
since February 2005. 

Example: 

1234 W35 -567 

 
 
 
 
 
 
 
 
 
 

Recommended test equipment and accessories 

Electrical safety tester (Example: Unimet 1000 ST, Bender) 

Electrosurgical analyzer (Example: QA-ES, Metron) 

 

Load resistor 100 

Ω, 25 W (or more), 5 % tolerance, low inductive part, short time load 

Connection cables, HF-output (banana) to electrosurgical analyzer or load resistor 

Power cord 

Footswitch 

 

 
 
 
 
 
 
 
 
 
 
 
 

Hardware version 

Summary of Contents for CelonLab ENT

Page 1: ...maintenance take care of the different hardware versions which may be applicable see below The hardware version of the unit can be identified by the serial number which can be found on the type plate...

Page 2: ...ountry medical device information Medizinproduktebuch 020 Labels present and legible Sample of symbol Unit is defibrillator safe Visible on the front panel see fig 2 Sample of symbol Caution Read the...

Page 3: ...nnection for equipotential bonding Visible on the back of the unit see fig 1 Samples of label for power fuses Visible on the back of the unit see fig 1 Sample of symbol Non ionizing radiation Visible...

Page 4: ...Celon AG Page 4 9 Issue date 01 09 2008 SCI 991007 1 4 Safety Check Instruction doc Fig 1 Sample of the backside of the CelonLab ENT power control unit Fig 2 Sample of the front panel of the CelonLab...

Page 5: ...Documentation 110 Document the test results in SCR 991007 3 Power output No Procedure Test equipment Electrosurgical analyzer Example QA ES Metron Connection cables HF output banana to the electrosurg...

Page 6: ...l analyzer according to the analyzer s instructions for use Set the power level of the unit to 12 and measure the following parameter Attention Start check procedure if not already done after setting...

Page 7: ...loaded and unloaded at rated load Connect the unit via the power cord with a rated supply voltage Connect the footswitch with the unit 280 Connect the load resistor 100 with the HF output 1 and 2 and...

Page 8: ...ety tester Example Unimet 1000 ST Bender Load resistor 100 25 W or more 5 tolerance low inductive part short time load Connection cables HF output banana to the load resistor Test accessories Power co...

Page 9: ...c 8 Inspection label No Procedure 430 Cover or exchange an inspection label as shown below at the enclosure back of the unit and mark the due date of the next periodic safety check month year The unit...