Chapter 3 Cleaning, Disinfection and Sterilization Procedures
33
EVIS EXERA GIF/CF/PCF TYPE 160 Series REPROCESSING MANUAL
Attach the water-resistant cap (MAJ-583 for CF-Q160ZL/I only)
• The zoom connector of the endoscope is not waterproof.
Before immersing the endoscope, always attach the
water-resistant cap.
• If the seal ring on the zoom connector is scratched, the zoom
connector may no longer be waterproof. Send the endoscope
immediately to Olympus for repairs.
• Never immerse the water-resistant cap unless it is attached
to the endoscope. Water remaining inside the water-resistant
cap can be transferred to and damage the zoom connector.
1.
Confirm that the seal ring on the zoom connector is not scratched (see
Figure 3.17).
2.
Push the water-resistant cap straight onto the zoom connector and turn it
clockwise (see Figure 3.18 (1) and (2)).
3.
Confirm that there is no gap between the zoom connector and the
water-resistant cap (see Figure 3.18 (3)).
Attach the water-resistant cap (MAJ-942 for CF-Q160DL/I only)
• The UPD scope connector of the endoscope is not
waterproof. Before immersing the endoscope, always attach
the water-resistant cap.
• If the seal ring on the UPD scope connector is scratched, the
UPD scope connector may no longer be waterproof. Send
the endoscope immediately to Olympus for repairs.
• Never immerse the water-resistant cap unless it is attached
to the endoscope. Water remaining inside the water-resistant
cap can be transferred to and damage the UPD scope
connector.
1.
Confirm that the seal ring on the UPD scope connector is not scratched (see
Figure 3.17).
2.
Push the water-resistant cap straight onto the UPD scope connector and
turn it clockwise (see Figure 3.18 (1) and (2)).
3.
Confirm that there is no gap between the UPD scope connector and the
water-resistant cap (see Figure 3.18 (3)).