DMTA048-01EN, Rev. C, February 2013
50 Chapter 4
Motor running and
the inlet tube is full of
water but there is no
water flow or low
water flow
The inlet tube or the
outlet tube has a kink
Remove the kink.
Water outlet holes
(for the scanner or
wedge) are
obstructed
Remove dirt from the water outlet
holes.
Water leaks on
scanner, wedge, and
tube connections
Verify that the tubes are fully
inserted.
Damaged or
punctured water
outlet tube
Replace the water outlet tube.
Contact your local Olympus NDT
Service Center for a replacement.
Dirt on inlet filter
Clean the inlet filter by following
the procedure in section 4.1
“Maintaining the Water Inlet
Tube’s Suction Filter” on page 43.
CFU03 or CFU05:
Valve on the pump
partially closed
Turn the pump valve counter
clockwise to allow more water
flow.
CFU-PWZ: Pump has
just been primed
Let the pump run for about 2
minutes.
Bad inlet tube
connection
Follow the connection procedure
in section 2.3.1 “Connecting the
Water Supply Tube” on page 29.
Table 4 Troubleshooting
(continued)
Problem Description
Possible Cause
Solution
Downloaded from
Summary of Contents for CFU03
Page 18: ...DMTA048 01EN Rev C February 2013 14 Introduction Downloaded from ManualsNet com search engine...
Page 26: ...DMTA048 01EN Rev C February 2013 22 Chapter 1 Downloaded from ManualsNet com search engine...
Page 42: ...DMTA048 01EN Rev C February 2013 38 Chapter 2 Downloaded from ManualsNet com search engine...
Page 46: ...DMTA048 01EN Rev C February 2013 42 Chapter 3 Downloaded from ManualsNet com search engine...
Page 64: ...DMTA048 01EN Rev C February 2013 60 Chapter 5 Downloaded from ManualsNet com search engine...
Page 72: ...DMTA048 01EN Rev C February 2013 68 Index Downloaded from ManualsNet com search engine...