CH10/CH20
D. REPAIR PROCEDURE
D-1
1. Centering of Optical System and Voltage Adjustment
1-1 Optical axis adjustment of revolving nosepiece
ž
Set the jigs as shown in the left figure.
ž
Turn on the power of the microscope
frame, loosen two SCREWS
‚
, and
adjust the position of the REVOLVING
NOSEPIECE SPRING
•
to meet the
standard.
Standard
Within 3 graduations
1-2 Illumination center adjustment
ž
Turn the KNOB
•
of condenser counter-
clockwise, and remove the condenser by
moving it downwards with the ADJUST-
MENT KNOB
‚
. Figure-1
ž
Turn on the power of the frame.
ž
Look through the collector lens from the
hole for mounting the objective.
ž
Make sure that the brightest filament seen
through the collector lens is in the center of
the hole for mounting the objective. Figure-
2 (When it is displaced, adjust the position of
the lamp holder.)
1-3 Condenser center adjustment
ž
Set the jig as shown in the left figure.
ž
Focus on the collector lens with the centering
telescope (KN0029).
ž
Loosen the three screws (AB3X8SA) securing the
stage, and adjust the position of the stage so that
the contour of the collector lens can meet the
standard.
Standard
Within 20%
CH20 frame
Standard eyepiece (KN0048; with adapter-1)
Standard tube (BKN0003)
Revolving
nosepiece
spring
(AD1905)
Standard objective
(KN0015)
Screw (AD2472)
Figure-2
Figure-1
Centering telescope
(KN0029)
Screw
(AB3X8SA)