background image

No.

Parts  name

Screw

Grease Adhesive

Remark

BINOCULAR  TUBE

(A)

(B)

(C)

(D)

(E)

CUK3X6SA, 4pcs. (*1)
KNW3SA, 4pcs. (*2)

3PUTB2X4SA, 3pcs. (*3)

ACU3X3SA, 2pc. (*5)

 DIOPTER RING

COVER

COVER

INDICATOR  PLATE

(G)

3PUTB2X4SA, 3pcs. (*4)

Attach it with double coated

adhesive tape.

SLEEVE ASS’Y

CUK3X6SB, 2pcs. (*8)

(H)

PRISM ASS’Y

(I)

PRISM ASS’Y

CUK3X8SA, 2pcs. (*9)

CUK3X8SA, 2pcs. (*10)

(J)

(K)

(L)

TORQUE  PLATE

AWU3X4SA, 2pcs. (*11)

SPRING WASHER

WASHER

(M)

MOUNT

(N)

RING

(O)

(P)

(Q)

(R)

(S)

PRISM  MOUNT-L

WASHER

PRISM  MOUNT-R

RING

SHAFT

OT3189

OT3189

OT3189

Tension:

*Procedure for mounting indicator plate(D):

1) Set the interpupillary distance to the minimum (48mm).

2) Affix the indicator plate(D) such that the indicator plate line may become parallel to the position

indicated by the round convex of helicoid ass’y(F) and the round convex of sleeve ass’y(G).

HELICOID ASS’Y

(F)

CUK3X6SB, 2pcs. (*7)

Refer to D-1.

Revolving axis:

Follow the adjustment procedure

on D-3.

Left/right optical  axis:
Follow the adjustment procedure

on D-3.

Push it in the arrow directions

during assembling.

Exit pupil center:

Follow the adjustment procedure

on D-3.

Same as the above

Apply  grease to the contacting

surface only.

Absolute optical axis:

Follow the adjustment procedure
on D-3.

*Tension adjustment (moving force of  interpupillary distance  ):

1) Loosen the two scews(*11) that fix the torque plate(J).

2) The tension is increased by turning the torque plate(J) clockwise. It is decreased by turnig

it counterclockwise. ( refer to the figure on the previous page.)

3) Measure the tension using a tension gauge.

Standard:  Tension 5 - 15N  (OT3223)

* In this manual, left ( right) sleeve indicates the left (right) side with the observed condition.

OT3189

standard   5-15N (OT3223)

C-6

CKX31/CKX41

                     C.  DISASSEMBLY AND ASSEMBLY PROCEDURES

 HU3X4SA, 1pc. (*6)

Summary of Contents for CKX31

Page 1: ...CULTURE MICROSCOPES CKX31 CKX41 REPAIR MANUAL SCIENTIFIC EQUIPMENT DIVISION MARKETING DEPARTMENT All rights reserved Reproduction in whole or in part without written permission is prohibited 2H...

Page 2: ...nuals because the explanation of these parts are ommited as a common repair skill In optical adjustment section parfocality adjustmet is mainly explained on D 6 and D 12 The parfocality is required to...

Page 3: ...ical unit D 14 F LUBRICANTS AND CHEMICALS 1 Listofgreases F 1 E JIGS AND TOOLS 2 List of adhesives F 1 D REPAIR PROCEDURE 1 Overview of optical adjustment D 1 2 Adjustment procedure CKX31 D 3 2 1 Prep...

Page 4: ...changing the objective when combined with IX2 SLP Applicable when the CKX31 CKX41 is combined with the IX2 SLP 3 Main switch and light intensity adjustment knob are located on the front of microscope...

Page 5: ...trast effectiveness can not be obtained In a combination of IX2 SL objective other than the above appicable objective is possible to use as long as the objective corresponds to PHL PHC PH1 PH2 2 Brigh...

Page 6: ...er Substage 70 L X180 W mm Auto X Y stage spot scan mountable 4 5 Illumination system Light source 6V30W halogen bulb 6V30WHAL Average life time of bulb 100 hours Lamp socket U LS30 3 2 Filter holder...

Page 7: ...5 Dimensions 364 370 5 486 6 194 5 380 5 473 2 364 370 5 515 5 195 393 1 473 7 1 CKX31 1 CKX41 UNIT mm 237 6 237 6 CKX31 CKX41 A OUTLINE OF PRODUCT A 4...

Page 8: ...LCACH40X2PH PH2 CPL10XRC LCACH20XRC LCACH40XRC 45KB 3 L U RFLT50 32ND25 CKX RFA w B G WH10X WH10X H 35WH10X PWH10X U CT30 WHB10X CT 5 WHB10X H CK40 CPG CK40 CPG30 IX2 GCP CKX41SF w U LS30 3 2 CKX31SF...

Page 9: ...on 0 1mm of the KN0048 Absolute optical axis 0 1mm or less on the image surface Combining the standard eyepiece KN0048 with Adapter 1 a microscope frame product the standard objective KN0041 read the...

Page 10: ...ness Thickness gauge OT1949 Part Item Standard Method Exit pupil center Within 10 of objective s exit pupil diameter Combining the centering telescope KN0023 an objective PlanC10X UIS observation tube...

Page 11: ...Refer to the above description method exit pupil center and check the optical axis and exit pupil center 1 Revolving axis 1 Standard eyepiece KN0048 with adapter 1 2 Objective PlanC10x 3 Test plate co...

Page 12: ...crew mounting part Inner screw threads Grease OT2144 Inner screw threads Grease OT2144 Screw it to G Change the position to adjust absolute optical axis Parfocality adjustnent 1 CKX31 frame b Selectio...

Page 13: ...er to C 15 ACU3X6SA 1pc 7 OT2144 Adjust absolute optical axis For parfocality adjustment refer to D 6 Relay lens part when adjusting parfocality remove the lens frame 1 a and insert the selected washe...

Page 14: ...ame A B 1 D 2 C E 3 4 F G 5 6 H I a b c d 7 e 8 f g Screw mounting part Adhesive OT1026 Inner screw threads Grease OT2144 Screwit to H Disassembly of Mount g C 3 CKX31 CKX41 C DISASSEMBLYANDASSEMBLY P...

Page 15: ...ical axis For parfocality adjustment refer Relay lens part I when adjusting parfocality remove the lens frame 1 a and insert the selected washer under it If necessary remove the lens frame 3 c from th...

Page 16: ...rease OT3189 Grease OT3189 Grease OT3189 Contacting surface Grease OT3189 Tension adjustment heavy light Prism ass y pushing direction Prism ass y pushing direction only Adhesive tape C 5 CKX31 CKX41...

Page 17: ...nvex of sleeve ass y G HELICOIDASS Y F CUK3X6SB 2pcs 7 Refer to D 1 Revolving axis Follow the adjustment procedure on D 3 Left right optical axis Follow the adjustment procedure on D 3 Push it in the...

Page 18: ...ension adjustment Adhesive tape Grease OT3189 Grease OT3189 Grease OT3189 Grease OT3189 Scale seal Contacting surface Prism ass y and inclined prism Pushing direction when assembling Making it even in...

Page 19: ...Same as the above Apply grease to the contacting surface only Absolute optical axis Follow the adjustment procedure on D 8 CUTB3X10SA 2pcs 3 BINOCULAR TUBE CUKK3X6SA 4pcs 4 I T U V FIXING PLATE PRISM...

Page 20: ...ng guide unit M when assembling it push it in arrow directions Grease OT2008 Grease OT2008 Adhesive tape 8 Adhesive OT1026 Grease OT2006 Grease OT2006 Grease OT2012 Grease OT2012 Grease OT2012 gears f...

Page 21: ...e of the shaft a Apply grease to the outer surface Apply grease to the outer surface While holding the coarse adj knob ass y screw the fine shaft mount into it Check that the gears moves smoothly Appl...

Page 22: ...ide M Arrange the rollers altenately on the upper side so that they face in same direction to lower side Adjustment guide mounting position Working force standard 0 2 0 5N without play or tightness Gu...

Page 23: ...without play or OT3225 tightness Check the sliding surface of wire guide If there is a dent on it replace the wire guide Grease applied part each groove attached wire guide sliding surface of wire gui...

Page 24: ...11 a screws for fixing the bottom plate Also remove the screws 2 b screws for fixing the circuit board 8 9 10 3 Rheostat ass y CN1 to CN2 CN3 CKX31 CKX41 C DISASSEMBLYANDASSEMBLY PROCEDURES C 13 7 Ele...

Page 25: ...ctor Connection to CN3 rheostat ass y connector Connection to CN2 lamp connector Voltage adjustment it is necessary to adjust it if the rheostat ass y or circuit board F is replaced individually Refer...

Page 26: ...5 6 W X Y CKX31 CKX41 C DISASSEMBLYANDASSEMBLY PROCEDURES C 15 8 Transmitted illuminator Adhesive tape Adhesive OT1026 Adhesive OT1026 Adhesive OT1026 Adhesive OT1026 Grease OT2010 Grease OT2010 slidi...

Page 27: ...R Attach it with double coated adhesive tape G H CUKSK3X8SA 2pcs 3 L CSK3X6SA 3pcs 4 PSK2X4SA 3pcs 5 OT1026 N OT1026 Apply adhesive to 2 points V RING SCREW SPRING BALL X Y Z OT1026 Mount it in the co...

Page 28: ...arfocality in the left sleeve After checking it in the left sleeve turn the helioid to focus on the specimen in KN0041 then loosen the screws 1 only move the diopter ring and fix it at the position wh...

Page 29: ...Cover Cover L Cover R Fig 2 Fig 2 4 Parfocality CKX41 In the same manner as CKX31 adjust the paforcality by inserting washer under the lens frame 1 For adjustment procedure refer to D 12 2 1 Left sle...

Page 30: ...40 repair manual Jigs See D 4 Procedure See D 8 4 2 4 3 to D 10 in CK30 40 repair manual Jigs See D 4 Procedure See D 11 5 2 to D 13 in CK30 40 repair manual Jigs See D 4 Procedure See D 14 6 2 in CK...

Page 31: ...OT1949 KN0048 Adapter 1 2 PlanC10X 3 Absolute optical axis 1 Standard eyepiece KN0048 Adapter 1 2 Standard objective KN0041 2 PlanC10X 5 Parfocality 1 Standard eyepiece KN0048 Adapter 1 3 Standard ob...

Page 32: ...t does not meet the standard adjust the parfocality referring to D 6 Standard 0 3mm within 3 graduations 2 2 Checking parfocality for CKX31 Necessary jigs 2 Standard eyepiece KN0048 Adapter 1 1 Focusi...

Page 33: ...selection 5 Assemble the parts in reverse order of disassembly 6 Set the jigs focus on the reticle of KN0048 and the specimen in KN0041 At this time read the difference from the index on the helicoid...

Page 34: ...is 1 9 mm when the total thickness 1 70mm of washers 0 7mm 1pc and 1mm 1 pc are inserted parfocality difference becomes 0 06mm reading value standard 1 0 Unit mm 0 05 0 1 0 3 0 5 0 7 1 0 1 1 1 70 0 06...

Page 35: ...D 10 Procedure See E 8 to E 9 in CH3 BI45 repair manual Jigs See D 9 and D 10 Procedure See E 10 to E 11 in CH3 BI45 repair manual Jigs See D 9 and D 10 Procedure See D 14 6 2 in CK30 40 repair manual...

Page 36: ...2 2 3mm adapter ring KC2040 3 Focusing telescope FT 36 or U FT 4 Standard objective KN0041 5 UIS standard observation tube BXKN001 4 Stage squareness 1 Stage for checking stage tilt alignment SKN0003...

Page 37: ...ality 1 Standard eyepiece KN0048 Adapter 1 3 Standard objective KN0041 2 Focusing telescope FT 36 or U FT KN0003 or cross micrometer OB M1 100SQ 3 Test plate concentric circles KN0003 or OB M1 100SQ c...

Page 38: ...Necessary jigs 2 Standard eyepiece KN0048 Adapter 1 and Adapter 2 5 Standard objective KN0041 1 Focusing telescope FT 36 or U FT Set the following jigs as illustrated below to check the parfocality 3...

Page 39: ...turn the relay lens ass y counterclockwise to remove it 2 1 2 Remove the screws 1 and remove the part of relay lens ass y and TR body ass y Screw AB3X14SA 1 Washer BNW3SA 4pcs each Screw ACU3X6SA 2 1...

Page 40: ...iffernce is 2 0 mm when the total thickness 1 05mm of washers 0 05mm 1pc and 1mm 1 pc are inserted parfocality difference becomes 0 03mm reading value standard 0 7 Unit mm Numeric character 1 or 2 in...

Page 41: ...ly procedures on C 13 to C 14 A A Power cord VR21 VR22 VR23 6V30W halogen bulb Lamp cable a Minimum voltage adjustment 1 Turn ON the power Turn the light Power switch Set the digital multimeter to CN2...

Page 42: ...tering telescope Standard objective Standard eyepiece including adapter 1 and 2 Thickness gauge Tension gauge 30N Tension gauge 3N Gauge for checking stage tilt alignment Test plate 5 100mm concentric...

Page 43: ...ight Silicone grease Grease medium Grease heavy Mo grease 2 List of adhesives No Description Page OT3111 OT1378 OT1026 OT1838 Cyanoacrylate adhesive Solvent based adhesive transparent Solvent based ad...

Reviews: