Using a nylon-bristle brush, scrub the shelf that 
contains the Olympus logo and part number.

Flush the interior of the handpiece with cool distilled, 
deionized, or potable water. Actuate the retention 
mechanism while flushing. Ensure water flows through 
all drainage ports.

Use 70% isopropyl alcohol to wipe and flush the 
handpiece, power cord, and plug. Flush the exterior to 
remove any remaining residue.

Dry completely with a lint-free cloth. Visibly inspect the entire handpiece for any remaining contamination and repeat 
the cleaning procedure if needed.

Actuate the blade retention mechanism.

3

7

4

5

6

Automated cleaning can be performed in an automatic 
washer-disinfector as an alternative to manual cleaning.

The handpiece is designed for sterilization by steam autoclave. 
Handpieces can be sterilized wrapped or unwrapped.

 Place the handpiece and cord, on the washer rack or 
place them in the approved tray

 

(Item EGMDHPTRAY). 

 The washer should be set to the Vario TD cycle. After 
automated cleaning, visually inspect the handpiece for 
any remaining residue and repeat the cleaning procedure 
if needed.

AUTOMATED CLEANING

STERILIZATION

Steam sterilization

Cycle

Gravity

prevacuum

Temperature

132° C

132° C

time

15 minutes

4 minutes

Drying Time

60 minutes

30 minutes

To prevent damage, the handpiece must be completely vacuum dried before storage.

STORAGE

HANDPIECE CLEANING – QUICK-REFERENCE GUIDE 

1

2

Note:

 To satisfy international, regional, governmental, or health-

care facility requirements, the handpiece may be processed at 
temperatures up to 137° C without causing damage.