Problem
Cause
Remedy
Page
The abnormal temperature error
occurs.
One of the following boards is installed
in the adjacent slot on the fan side of
the PCIe interface board.
· Graphic board
· PCIe interface board of DP80, DP73
or DP72
· Board with power consumption of
3W or more
Move the relevant board to the different
slot.
6,
26
The image is too bright.
The exposure correction is set to the +
side beyond the appropriate value.
Return the exposure correction value
to 0 and set the appropriate exposure
correction value.
(Online
manual)
The dark area of the image is set as
the metering area.
Set the area where you want to meter
as the metering area.
(Online
manual)
AE lock, which was set when the
exposure time was longer than the
currently required exposure time, is
active.
Cancel the AE lock.
(Online
manual)
The input highlight level of the level
adjustment is too low.
Reset the current level adjustment and
adjust the level properly again.
(Online
manual)
The illumination of the microscope is
too bright.
Adjust the brightness by reducing
the illumination light intensity of
the microscope or engaging the
attenuation filter in the light path.
–