118
EMC information
CLV-190 INSTRUCTION MANUAL
App.
Guidance and manufacturer’s declaration — Electromagnetic
immunity
This model is intended for use by medical personnel in hospital environments and for use in the
electromagnetic environment specified below. The customer or the user of this model should
assure that it is used in such an environment.
This instrument can be used with the high-frequency electrosurgical equipment that designated
by Olympus.
Immunity test
IEC 60601-1-2
(2014)
test level
IEC 60601-1-2
(2007, 2001)
test level
Compliance
level
IEC 60601-1-2 (2007, 2001)
Electromagnetic
environment — Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
Contact:
±
8 kV
Air:
±
2,
±
4,
±
8,
±
15 kV
Contact:
±
2,
±
4,
±
6 kV
Air:
±
2,
±
4,
±
8 kV
Same as left
Floors should be made of wood,
concrete, or ceramic tile that
hardly produces static. If floors
are covered with synthetic
material that tends to produce
static, the relative humidity
should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±
2 kV
for power supply lines
±
1 kV
for input/output lines
±
2 kV
for power supply
lines
±
1 kV
for input/output
lines
Same as left
Mains power quality should be
that of a typical commercial
(original condition feeding the
facilities) or hospital
environment.
Surge
IEC 61000-4-5
Differential mode:
±
0.5,
±
1 kV
Common mode:
±
0.5,
±
1,
±
2 kV
for signal input/
output lines:
±
2 kV
Differential mode:
±
0.5,
±
1 kV
Common mode:
±
0.5,
±
1,
±
2 kV
Same as left
Mains power quality should be
that of a typical commercial or
hospital environment.
Summary of Contents for EVIS EXERA III
Page 2: ......
Page 6: ...iv Contents CLV 190 INSTRUCTION MANUAL...
Page 20: ...14 Summary of the Equipment Functions CLV 190 INSTRUCTION MANUAL...
Page 22: ...16 1 1 Checking the package contents list CLV 190 INSTRUCTION MANUAL Ch 1...
Page 32: ...26 2 3 Rear and side panels CLV 190 INSTRUCTION MANUAL Ch 2...
Page 98: ...92 6 4 Lamp usage indicator reset CLV 190 INSTRUCTION MANUAL Ch 6...
Page 108: ...102 7 8 Transportation CLV 190 INSTRUCTION MANUAL Ch 7...