4.12 Inspection of the high intensity mode
59
CLV-190 INSTRUCTION MANUAL
Ch.4
NOTE
Pressing any lit button except the lamp button and counter reset button on the
control panel cancels the transillumination function.
NOTE
• The high intensity mode is not available when no endoscope is connected to the
light source.
• New products released after the introduction of the light source may also be
compatible for use in combination with the light source. For further details, refer to
the instruction manual for the endoscope or the light guide cable to be used or
contact Olympus.
2
Confirm that the transillumination button blinks,
and the brightness indicator indicates the
maximum level (see Figure 4.17). The
examination light changes to the maximum
intensity for about 7 seconds.
Figure 4.18
3
Confirm that the examination light changes to its original brightness after 7 seconds.
4.12 Inspection of the high intensity mode
Distal end of
endoscope