4.13 Inspection of air and water feeding
63
CLV-190 INSTRUCTION MANUAL
Ch.4
7
Release the hole of the air/water valve of the endoscope.
8
Remove the distal end of the endoscope from the sterile water.
9
Depress the air/water valve of the endoscope.
10
Press the airflow regulator button to change the airflow level setting and confirm that
the amount of water exiting from the air/water nozzle changes accordingly.
11
Release the air/water valve of the endoscope.