Other
Issue
Cause
Solution
The remote switch does not work. The wrong remote switch is operated.
Operate the correct remote switch.
The remote switch function has been
set incorrectly.
Set the remote switch function correctly
as described in the video system centre’s
instruction manual.
For BF-Q190 and BF-H190-The remote
switch is left depressed due to a
difference between internal and
external pressures of the endoscope
that is made in the pressure reduction
process of gas sterilization.
If the sterilization cap is attached to the
venting connector, detach it. If not, attach
the sterilization cap to the venting
connector once and then detach it.
The remote switch 1 does not
return to the OFF position.
Remote switch 1 is pressed strongly
from the side.
Pull it upwards gently.
For the TJF-Q180V endoscopes-
Guidewire cannot be locked.
Guidewire is not locked at its stiff part.
Lock the guidewire at its stiff part.
Guidewire with a diameter less than ø
0.64 mm is used.
Select a guidewire with a diameter of ø
0.64 mm or more.
Guide wire-locking groove is dirty.
Clean and disinfect or sterilize the
Guide wire-locking groove as described
in the “REPROCESSING MANUAL” with
your endoscope model listed on the
cover.
Contrast media is congealed in the
guide wire lumen of the EndoTherapy
accessory.
Clean the lumen of the EndoTherapy
accessory and then insert/withdraw it.
No light or air is emitted from the
end of the scope.
the endoscope is not connected to the
output socket
Corrosion of electrical contacts or
shortage of light control contacts
ensure endoscope is completely dry
before inserting into connector socket
Poor Airflow through light source
Collection of dust in air intake located at
base of unit. Can cause reduction of
lamp life.
Ensure unit is free of dust and debris
Error Codes: Here are some examples of error codes
(taken from the CV-190 manual)
Code Error Message
Cause
Solution
A27
CLV is not connected (A27)
The light source is connected
incorrectly, or a wire in the
light source cable is
disconnected
Connect it correctly as described in Section
3.5, “Connection of the light source”.(CV-190
manual)
E102
CLV lamp ERR E102
Please check examination
lamp
Patient data out of the input
Immediately turn OFF the light source and
inspect the lamp as described in the
instruction manual for the light source. If no
irregularity is observed on the lamp,
immediately turn OFF the light source, unplug
the power cord from the wall mains outlet and
the AC power inlet on the light source, and
then contact Olympus.
E204
Focus function ERR E204
Contact Olympus
The focus function of the
endoscope or the video
system center does not work.
Stop using the endoscope and withdraw it
from the patient as described in the instruction
manual for the endoscope. Also, contact
Olympus. This message can be hidden
temporarily by pressing the “ESC” key
on the keyboard
E302
Internal buffer FULL E302
Please transfer unsaved
images now
The internal buffer is full of
image data that has not been
transferred.
Transfer the data as described in CV-190
manual
E303
Portable memory format
ERR
The portable memory cannot
be formatted.
Format the portable memory again as
described in CV-190 Manual