Chapter 3 Cleaning, Disinfection and Sterilization Procedures
27
EVIS LUCERA JF/TJF TYPE 260V REPROCESSING MANUAL
6.
Detach the AW channel cleaning adapter, washing tube, suction valve and
biopsy valve from the endoscope and place them in a container of detergent
solution.
Detach the distal cover
• When detaching the distal cover, forcefully grasping other
parts of the bending section can result in damage to the
bending mechanism of the bending section or deform its
covering.
• After the distal cover has been removed, make sure not to
subject the forceps elevator to impact or bend the elevator
wire during cleaning, disinfection, sterilization, storage or
preparation of the endoscope. This could adversely affect the
function of the forceps elevator.
1.
Move the elevator control lever all the way in the opposite direction of the
“
U” direction until it stops.
2.
Gently hold the distal part in the covering of the bending section (see Figure
3.15). Hold the top end of the distal cover with the other hand.
Figure 3.15
Distal part