Chapter 3 Cleaning, Disinfection and Sterilization Procedures
41
EVIS LUCERA JF/TJF TYPE 260V REPROCESSING MANUAL
Figure 3.26
Flushing the interior of the forceps elevator
1.
Immerse the distal end in detergent solution, and operate the elevator
control lever to raise and lower the forceps elevator 3 times.
2.
With the forceps elevator raised, flush the interior of the forceps elevator
with detergent solution using the 30 cm
3
(30 ml) syringe (see Figure 3.27).
Figure 3.27
Weighted end
Instrument
channel port
Suction channel
Suction cleaning adapter
Insertion tube
Instrument
channel
Suction
cylinder
Universal cord
Endoscope connector
Suction connector
Suction tube
Suction pump
Suction bottle
Forceps elevator
Detergent solution