46
Chapter 3 Cleaning, Disinfection and Sterilization Procedures
EVIS LUCERA JF/TJF TYPE 260V REPROCESSING MANUAL
8.
Using the 30 cm
3
(30 ml) syringe, flush the air/water and suction channels
with 90 cm
3
(90 ml) of air, respectively, via the injection tube (see Figures
3.30 and 3.33).
9.
Using the 5 cm
3
(5 ml) syringe, flush the elevator wire channel with 10 cm
3
(10 ml) of air via the washing tube (see Figures 3.31 and 3.32).
10.
Remove the clean, lint-free cloth from the distal end and the control section
of the endoscope.
11.
Detach the channel plug and injection tube from the endoscope.
12.
Disconnect the washing tube from the endoscope.
Figure 3.33
Dry external surfaces
1.
Using a clean lint-free cloth, wipe and dry the external surfaces of the
endoscope and all equipment thoroughly.
2.
Inspect the endoscope and all equipment for residual debris. Should debris
remain, repeat the procedures given in this section.
Suction channel
Biopsy valve cap
Channel plug
Suction cylinder
Suction connector
Suction port
Injection tube
Suction channel port
Air/water channel port
Syringe
Suction channel