56
Chapter 3 Cleaning, Disinfection and Sterilization Procedures
EVIS LUCERA JF/TJF TYPE 260V REPROCESSING MANUAL
3.9
Cleaning, disinfection and sterilization
procedures for reusable parts and reprocessing
equipment
This section includes the cleaning, disinfection and sterilization procedures for
the reusable parts and reprocessing equipment listed below.
Equipment needed
Prepare the following equipment, and wear appropriate personal protective
equipment.
• Air/water valve
(MH-438)
• Suction valve
(MH-443)
• AW channel cleaning adapter
(MH-948)
• Mouthpiece
(MB-142)
• Biopsy valve
(MB-358)
• Channel-opening cleaning brush
(MH-507)
• Channel cleaning brush
(BW-20T)
• Suction cleaning adapter
(MH-856)
• Sterile lint-free cloths
• Sterile water for sterile water rinse
• Clean water (non-sterile water)
• 70% ethyl or isopropyl alcohol
• Clean lint-free cloths
• Personal protective equipment
• Detergent solution
• Soft brush
• Disinfectant solution
• Small basins
• 30 cm
3
(30 ml) syringe
• Channel cleaning brush
(BW-20T)
• Channel-opening cleaning brush
(MH-507)