12
Chapter 2 Compatible Reprocessing Methods and Chemical Agents
EVIS LUCERA JF/TJF TYPE 260V REPROCESSING MANUAL
• Exceeding the recommended parameters may cause
equipment damage (see Table 2.3).
Table 2.3
Steam sterilization (autoclaving) exposure parameters
Process
Parameters
Prevacuum
Temperature
132 – 134°C
(270 – 274°F)
Exposure time
5 minutes